Programmable measurement and control system: BlueBox TS
Operating system:
Embedded Linux
Version: selectable option
Available operating system languages: English, German, French and Spanish, others on request.
Memory size: selectable option
Circular buffer for programs and measured data
Description:
Touch panel, LCD colour display 480×272 pixel with 65.536 colours
All in- and outputs are pluggable (exception fibre optic cable)
Aluminium housing (LxWxH): 280 mm x 90 mm x 170 mm, IP 65
Colour: blue powder-coated RAL 5010
Weight: 2.9 kg
Storage temperature range: -10 °C … +50 °C
Operating temperature range: -20 °C … +45 °C
Power supply: 24 V DC, 3.2 A (The power supply unit is not included in the delivery!)
The system is designed for the connection of up to 200 sensors and actuators.
Number of virtual sensors: selectable option
Interfaces:
1x Serial Interface (example protocols: RS232/RS485 (Modbus RTU), IEC 60870-5-104; further more, see protocol list)
1x CAN
1x Ethernet 10/100MBit
1x USB
2x pulse input
2x 4-20mA output
2x relay output 48V AC/DC 0,5A
1x 12V 6W integrated power supply for sensors with RS232 / RS485 port
Optional GPRS/UMTS module (not included in the delivery)
Device concept:
Purely optical sensor head with optical fibre connection
Remotely transferable calibration algorithms for parameter calculation
Available comprehensive software for performing calibrations
All calibration settings are accessible, accessible raw spectra
Freely adjustable path length from 0.5 to 20 mm
ISA – In situ spectral analyser – ATEX category 3
ISA is an In situ submersible spectrum analyser.
Measurements: UV/VIS spectra in range of 200 – 710 nm
Resolution: 256 Pixel
Measuring interval min.: 10 sec.
Light source: Xenon flash light
Automatic cleaning by air flushing and valve for connection to a compressed air line are included in the delivery.
Measurement head:
Material: stainless steel (1.4404), optionally titanium
Dimensions (sensor head):
length: approx. 230 mm (dependent on the path length), Diameter: 44 mm
Weight: approx. 1.5 kg
Temperature range: 0 °C … +110 °C
Degree of protection: IP68, ATEX category 3 (optionally category 2)
Length of the fibre optic cable: selectable option
Areas of application: process control, waste water monitoring, water quality and environmental monitoring, etc.
Multi – parameter calculation (max. 50)
Included parameters on delivery:
SAC, Fingerprint, Spectral alarm
Selectable options | Version | Memory size |
Number of virtual sensors |
Length of fibre optic cable |
|
Version |
|||||
Pos. | |||||
1 | Version 4.2 | 42GO | |||
Memory size |
|||||
Pos. | |||||
3 | 512 MB | -512 | |||
4 | 2GB | -2G | |||
Number of virtual sensors |
|||||
Pos. | |||||
5 | 10 | -10 | |||
6 | 20 | -20 | |||
7 | 50 | -50 | |||
8 | 100 | -100 | |||
Length of fibre optic cable |
|||||
Pos. | |||||
9 | 2,5 m | -2,5 | |||
10 | 6 m | -6 | |||
11 | 10 m | -10 | |||
12 | 15 m | -15 | |||
13 | 20 m | -20 | |||
14 | 6 m with titanium measuring head | -6T | |||
Example configuration Art. no. 486 0060-41GO-512-10-2,5 |