BlueMon Total Nitrogen & Total Phosphorus

Fully automatic on-line analyser for measuring concentrations of mediums
in fluids according to wet-chemical methods.
The photometer total nitrogen & total phosphorus version of the BlueMon system comprises of the following equipment: analyser unit and control and measurement electronics in a powder-coated cabinet.

Measuring cell: photometer
(wavelength and measuring gap is dependent on the method)

Measuring range: selectable option

The choice of procedure determines the number of pumps, valves and
the photometer configuration with regard to wavelength and cuvette.

Number of sample channels: selectable option
Number of free valves: dependent on procedure
e.g. use as dilution channel (e.g. H2O)
Number of peristaltic pumps: dependent on procedure
Number of process valves: dependent on procedure

Technical details:
Protection classification: IP54 (optional IP65)
Dimensions: (LxWxH): 600 mm x 300 mm x 700 mm
Cabinet material: aluminium (powder-coated)
Colour: blue (RAL 5010)
Sample pressure: 0 bar (max 0.05 bar overpressure)
Sample temperature: 10 °C … 35 °C
Sample flow rate: 2 … 10 l/h, no suspended solids
Samples with particle size > 30 microns need filtration
Environmental temperature: 15 °C … 35 °C
Operating system: embedded Linux
Power consumption (average): 45W

External power supply unit:
Input voltage range: 85…265V AC
Output: 230 V/50 Hz (power supply included)

Measurement, control inputs and outputs
Interfaces:
1x Ethernet (incl. Modbus)
1x RS-232 or RS-485 (incl. Modbus)
1x CAN bus (for connecting further BlueBox System modules for sensors and actuators)
2x current output 4 – 20mA
Additional interfaces: selectable option

Sensors:
1x photometer interface with power adjustment for brightness control
1x connector for pH glass electrode
1x connector for temperature (PT1000) 0 – 80 °C
1x connector redox electrode
1x current input 4 – 20mA
6x digital in
1x connector for leakage sensor
6x connector bubble detector for sample reagent

Actuators:
1x output for digester with temperature, heating (0-150 °C) and UV monitoring
1x control stirrer
2x motor control forward/backward and pulse counter
5x motor control – (direction jumper selectable)
12x relays for valves, limit and interference signals

Selectable optionsMeasuring
range
Number
of sample
channels
Additional
interfaces
Chemical
drainage
 Measuring range
 
Pos.     
1Total nitrogen up to 6 mg/l
Total phosphorus up to 1 mg/l
61
   
2Total nitrogen up to 100 mg/l
Total phosphorus up to 20 mg/l
120   
 
Number of sample channels
 
Pos.    
22 channels-/  
 
Additional interfaces
 
Pos.   
3Without-/ 
42x 4-20 mA-2mA 
54x 4-20 mA-4mA 
      

Leave a Reply