BlueMon Photometer Orthophosphate

Fully automatic on-line analyser for measuring concentrations of mediumsin fluids according to wet chemical methods.
The photometer orthophosphate version of the BlueMon system comprises of the following equipment: analyser unit and control and measurement electronics in a powder-coated cabinet.

Measuring cell: photometer
(wavelength and measuring gap is dependent on the method)

Method: analysis of the orthophosphates with molybdenum blue method
Wavelength: 810 nm
Measuring range: selectable option

The choice of procedure determines the number of pumps, valves and
the photometer configuration with regard to wavelength and cuvette.

Number of sample channels: selectable option
Number of free valves: dependent on procedure
e.g. use as dilution channel (e.g. H2O)
Number of peristaltic pumps: dependent on procedure
Number of process valves: dependent on procedure

Technical details:
Protection classification: IP54 (optional IP65)
Dimensions: (LxWxH): 450 mm x 260 mm x 450 mm
Cabinet material: aluminium (powder-coated)
Colour: blue (RAL 5010)
Sample pressure: 0 bar (max 0.05 bar overpressure)
Sample temperature: 10 °C … 35 °C
Sample flow rate: 2 … 10 l/h, no suspended solids
Samples with particle size > 30 microns need filtration
Environmental temperature: 15 °C … 35 °C
Operating system: embedded Linux
Power consumption (average): 45W

External power supply unit:
Input voltage range: 90…264V AC
Output: 24 V DC, 10 A, 240 W (power supply included)

Measurement, control inputs and outputs
Interfaces:
1x Ethernet (incl. Modbus)
1x Serial Interface (example protocols: RS232/RS485 (Modbus RTU), IEC 60870-5-104; further more, see protocol list)
1x CAN bus (for connecting further BlueBox System modules for sensors and actuators)
2x current output 4 – 20mA
Additional interfaces: selectable option

Sensors:
1x photometer interface with power adjustment for brightness control
1x connector for pH glass electrode
1x connector for temperature (PT1000) 0 – 80 °C
1x connector redox electrode
1x current input 4 – 20mA
2x digital in
1x connector for leakage sensor
2x connector bubble detector for sample reagent

Actuators:
1x output for digester with temperature, heating (0-150 °C) and UV monitoring
1x control stirrer
1x motor control forward/backward and pulse counter
2x motor control – (direction jumper selectable)
6x relays for valves, limit and interference signals

Selectable options Measuring
range
Number
of sample
channels
Additional
interfaces
 Measuring range
Pos. Accuracy   
10,0 – 50 mg/l Orthophosphate3%50d  
 
Number of sample channels
Pos.   
22 channels-C2 
33 channels-C3 
44 channels-C4
 
 
Additional interfaces
Pos.  
5Without-/
62x 4-20 mA-2mA
74x 4-20 mA-4mA
      
 Example configuration Art. no. 488 1FP0-50d-C2-/